About Us 

 

 

 Products  Sales Contacts Support  |  Service/Calibration  News

 Home > Products > Sound Level Meters > SoundTrack LxT > Instrument Software > Time History

SoundTrack LxT  Software Overview

 

..................................................

SoundTrack LxT Integrating SLM

 

..................................................

SoundTrack LxT Data Displays

 

..................................................

Soundtrack LxT Time History (LOG)

 

..................................................

SoundTrack LxT 1/1 Octave Band Analyzer (OB1)

 

..................................................

SoundTrack LxT 1/1 and 1/3 Octave Band Displays (OB3)

 

..................................................

SoundTrack LxT Digital Voice Annotation (DVA)

 

..................................................

SoundTrack 120MB Memory (MEM120)

 

..................................................

 

Documentation

SoundTrack LxT - Instrument Software PDF

SoundTrack LxT - Feature Chart PDF

 

 

 

 
 

SoundTrack LxT - Time History

 

With the optional Data Logging firmware installed, when in the Time History mode the SoundTrack LxT can automatically log up to 16 acoustic and non-acoustic parameters, at time intervals from 1 second to 24 hours. The control of the time intervals is achieved via the following 6 Run Control modes of the instrument:

  • Manual Stop: initiated and stopped manually

  • Timed Stop: initiated manually and automatically stopped after a user-defined period.

  • Stop When Stable: initiated manually and automatically stopped after a user-defined period and when the measured level has remained within a user-defines band.

  • Continuous: initiated and stopped manually as in the Manual Stop mode. The exception being that the Continuous mode has a user-selected capability of a 24 hour, user time set, auto-store.

  • Single Block Timer: initiated and stopped manually, when measurement have been made in the time interval defined by the single block timer.

  • Daily Timer: initiated and stopped manually, when measurement have been made during the separate time blocks defined in the instrument setup.

The 16 measurement parameters are:

  • RMS Levels: Leq; Lmax; Lmin (with predefined A/C/Z frequency weightings and Slow/Fast/Impulse detectors)

  • Peak Level: Lpeak (with predefined A/C/Z frequency weighting)

  • Specialized Acoustic Parameters:

           - LAFTM5 Tacktmaximal 5 (used in Germany)

           - Ltwa1 and Ltwa2 time-weighted averages (associated with

             Dose 1and Dose 2 exchange rates and thresholds for

             Dosimeters 1 & 2

           - LCSeq - LASeq

  • 1/1 and 1/3 Octave Spectra: the following frequency spectra can be logged as 1/1 Octave Bands (OB1 firmware) or 1/1 & 1/3 Octave Bands (OB3 firmware):

           - OBA 1/1 Leq

           - OBA 1/1 Lmax

           - OBA 1/1 Lmin

           - OBA 1/3 Leq

           - OBA 1/3 Lmax

           - OBA 1/3 Lmin

  • Non-Acoustic Parameters: (from LxT)

           - Battery Level

           - Internal Temperature

 

Further Information ...

 

(c) 2007 PCB Group, Inc. In the interest of constant product improvement, specifications are subject to change without notice. PCB, IMI with associated logo, ICP, Modally Tuned, Swiveler, Spindler, TORKDISC, Spark, Blaze, and SoundTrack LxT are registered trademarks of PCB Group, Inc. SensorLine is a service mark of PCB Group, Inc. All other trademarks are properties of their respective owners.